Good news! The PRISM website is available for submissions. The planned data migration to the Scholaris server has been successfully completed. We’d love to hear your feedback at openservices@ucalgary.libanswers.com
 

Safety of in-hospital insertable cardiac monitor procedures performed outside the traditional settings: results from the Reveal LINQ in-office 2 international study

Abstract

Abstract

            Background
            Historically, the majority of insertable cardiac monitor (ICM) procedures were performed in the cardiac catheterization (cath) lab, electrophysiology (EP) lab, or operating room (OR). The miniaturization of ICMs allows the procedure to be relocated within the hospital without compromising patient safety. We sought to estimate the rate of untoward events associated with procedures performed within the hospital but outside the traditional settings and to characterize resource utilization, procedure time intervals, and physician experience.
          
          
            Methods
            The Reveal LINQ in-Office 2 (RIO 2) International study was a single arm, multicenter, prospective study. Patients indicated for an ICM and willing to undergo device insertion outside the cath/EP lab or OR were eligible and followed for 90 days after insertion.
          
          
            Results
            A total of 191 patients (45.5% female aged 63.8 ± 26.9 years) underwent successful Reveal LINQ ICM insertion at 17 centers in Europe, Canada and Australia. The median total visit duration was 106 min (interquartile range [IQR]: 55–61). Patient preparation and patient education accounted for 10 min (IQR: 5–20) and 10 min (IQR: 8–15) of total visit duration, respectively. Preparation and education occurred in the procedure room for 90.6 and 60.2% of patients, respectively. There were no untoward events (0.0, 95% CI: 0.0–2.1%) though four patients presented with procedure-related adverse events that did not require invasive intervention. Physicians rated procedure location as convenient or very convenient.
          
          
            Conclusions
            The Reveal LINQ™ ICM insertion can be safely and efficiently performed in the hospital outside the cath/EP lab or OR.
          
          
            Trial registration
            ClinicalTrials.gov identifier 
                NCT02412488
                
              ; registered on April 9, 2015.

Description

Keywords

Citation

BMC Cardiovascular Disorders. 2019 May 31;19(1):132